Your browser doesn't support javascript.
loading
Investigation on shielding reconstruction with clay bricks for medical linear accelerator room / 医疗卫生装备
Chinese Medical Equipment Journal ; (6): 81-84,88, 2017.
Artículo en Chino | WPRIM | ID: wpr-662250
ABSTRACT
Objective To provide a solution to improve the radiation shielding for medical accelerator room with clay bricks.Methods According to the basic shielding equations provided by the relevant national standards and international reports,equations for shielding reconstruction calculation were derived to calculate the required thickness of clay bricks wall,which reached the reference dose rate control level.The dose rates at reference points after reconstruction were also estimated based on the measured dose rates at reference points before reconstruction of accelerator room,calculated distance correction factors and transmission factors of added clay bricks wall.Results The measured dose rates satisfied the requirement of reference dose rate control level.The estimated dose rates were slightly higher than the measured values.Conclusion The radiation shielding requirement of medical accelerator room is fulfilled with the presented shielding reconstruction method by building clay bricks wall,measured dose rates could be conservatively predicted with the dose rates calculated by derived equations.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Medical Equipment Journal Año: 2017 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio pronóstico Idioma: Chino Revista: Chinese Medical Equipment Journal Año: 2017 Tipo del documento: Artículo