Fabrication of multiwalled carbon nanotube-surfactant modified sensor for the direct determination of toxic drug 4-aminoantipyrine☆ / 药物分析学报
Journal of Pharmaceutical Analysis ; (6): 231-238, 2015.
Artículo en Chino | WPRIM | ID: wpr-672175
ABSTRACT
A multi-walled carbon nanotube (MWCNT)-cetyltrimethylammonium bromide (CTAB) surfactant composite modified glassy carbon electrode (GCE) was developed as a novel system for the determination of 4-aminoantipyrine(AAP). The oxidation process was irreversible over the pH range studied and exhibited a diffusion controlled behavior. All experimental parameters were optimized. The combination of MWCNT-CTAB endows the biosensor with large surface area, good biological compatibility, electricity and stability, high selectivity and sensitivity. MWCNT-CTAB /GCE electrode gave a linear response for AAP from 5.0 × 10-9 to 4.0 × 10-8 M with a detection limit of 1.63 × 10-10 M. The modified electrode showed good selectivity against interfering species and also exhibited good reproducibility. The present electrochemical sensor based on the MWCNT-CTAB/GCE electrode was applied to the determination of AAP in real samples.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Journal of Pharmaceutical Analysis Año: 2015 Tipo del documento: Artículo