Comparison of routine test kits of creatinine in human serum with reversed-phase high-pressure liquid chromatographic assay / 中华检验医学杂志
Chinese Journal of Laboratory Medicine ; (12)2000.
Artículo en Chino | WPRIM | ID: wpr-685448
ABSTRACT
98.95%.The results of certified reference materials were consistent with the target value,and average deviation was -0.31%~1.35%.HPLC was served as the independent variable.When ereatinine was 200 ?mol/L,Bias of Beckman LX20 system,Vitros dry chemistry system and enzymatic method were -10%~13%,-13%~14% and -20%~10%,respectively.Bias of enzymatic method results was mostly negative,when creatinine was

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Laboratory Medicine Año: 2000 Tipo del documento: Artículo