ABSTRACT
A method for the determination of S-Adenosyl-L-methionine in fermentation liquid using the high performance liquid chromatography coupled with UV detector (HPLC-UV) was developed.A good linearity was obtained in the range of 0.1~1.0g/L (r = 0.9969) for S-Adenosyl-L-methionine.The average recoveries of S-Adenosyl-L-methionine were 99.89%~101.7% and the relative standard deviations were 0.48%~1.36%.The method is simple and rapid with reproducibility for the determination of S-Adenosyl-L-methionine in ferment liquid.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Tipo de estudio: Estudio diagnóstico Idioma: Chino Revista: China Biotechnology Año: 2006 Tipo del documento: Artículo