Calibration and quality control of medical image monitor / 医疗卫生装备
Chinese Medical Equipment Journal
; (6): 88-90,94, 2017.
Article
en Zh
| WPRIM
| ID: wpr-699909
Biblioteca responsable:
WPRO
ABSTRACT
Objective To explore the standard and methods for the calibration of the medical image monitor,and to execute its quality detection.Methods The related standards and concepts of medical image monitor were analyzed,and the grayscale standard display function (GSDF) defined by DICOM was discussed.Calibration of medical image monitor was implemented with the 14th part of DICOM standard on medical image display and DICOM GSDF standard function.Results It's pointed out that lookup table and brillmeter were indispensible for DICOM GSDF calibration of medical image monitor.The calibration modes included the ones of manual,automatic and network.Conclusion Medical image monitor has to undergo calibration and quality control conventionally,and DICOM GSDF calibration should be executed before its clinical application.