Research on energy respond characteristics of new diagnostic dosimeter / 中国医学装备
China Medical Equipment ; (12): 58-60, 2018.
Article en Zh | WPRIM | ID: wpr-706549
Biblioteca responsable: WPRO
ABSTRACT
Objective: To carry out comparative study on energy respond characteristics of 4 kinds of medical diagnostic X-ray dosimeters that included general new imported equipment and domestic equipment.Methods: In the standard radiation field of secondary standard dosimetry laboratory(SSDL)of IAEA,substitution method was adopted to complete contrast measurement of 10 kinds of X-ray diagnosis radiations which voltage were 60-150 kV,tube current were 10 mA and 2.5 mmAl was used as filterable addition.Results: The results of energy respond performances of domestic RD-98 diagnosis dosimeter and imported dosimeter indicated that all of these responses of different energy to dose were within ±5%when X-ray range was 6-150 kV.Conclusion: The energy response characteristics of 4 kinds of new diagnosis X-ray dosimeter were better,and the response of different energy to dose conforms to the requirement of IEC in diagnosis dosimeters of semiconductor detector and air ionization chamber.
Palabras clave
Texto completo: 1 Índice: WPRIM Tipo de estudio: Diagnostic_studies Idioma: Zh Revista: China Medical Equipment Año: 2018 Tipo del documento: Article