Analysis of Myocardial Function Using Gated Myocardial SPECT : Comparison of QGS, 4D-MSPECT Software and Echocardiography / 핵의학분자영상
Nuclear Medicine and Molecular Imaging
;
: 435-443, 2008.
Artículo
en Coreano
| WPRIM
| ID: wpr-71160
ABSTRACT
PURPOSE:
Gated myocardial perfusion SPECT provides not only myocardial perfusion status but also various functional parameters of left ventricle. We compared left ventricular ejection fraction, end-diastolic volume, LV mass by cardiac SPECT using Quantitative Gated SPECT (QGS), 4D-MSPECT software and standard 2D-echocardiography. MATERIALS ANDMETHODS:
One hundred fourteen patients (male 51, female 63; 29-85 years old, mean 61.3+/-13.3 years old) with normal perfusion status on Tc-99m tetrofosmin gated myocardial perfusion SPECT were analyzed retrospectively. Ejection fraction (LVEF), End-diastolic volume (LVED), LV mass (LVM) were calculated using QGS, 4D-MSPECT, and LVEF, LVM using 2D-echocardiography. Statistical analysis including Bland-Altman plot was performed using MedCalc(R) (MedCalc software, Mariakerke, Belgium).RESULTS:
The correlation of LVEF between methods was good 0.95/0.96 (stress/rest) between QGS and 4D-MSPECT, 0.79 between QGS and echocardiography, 0.79 between 4D-MSPECT and echocardiography (p<0.001). Using Bland-Altman plot, the 95% confidence interval of agreement between QGS and 4D-MSPECT ranged from -12.7% to 7.3% / from -12.2% to 6.5% (stress/rest). The agreement between QGS and echocardiography, 4D-MSPECT and echocardiography ranged from -17.4% to 24.0%, and -14.8% to 27.0% respectively. The correlation of LVM between methods was also good 0.95 between QGS and 4D-MSPECT, 0.76 between QGS and echocardiography, 0.73 between 4D-MSPECT and echocardiography (p<0.001). The 95% confidence interval of agreement between QGS and 4D-MSPECT ranged from -33.8 g to 14.1 g (stress/rest). The 95% confidence interval of agreement between QGS and echocardiography, 4D-MSPECT and echocardiography ranged from -148.7 g to 21.8. g, and -142.8 g to 35.5 g, respectively.CONCLUSION:
There was a good correlation for LVEF, LVED, LVM among methods (QGS, 4D-MSPECT, echocardiography), but the variance between methods was big. Therefore, the functional parameters by each method cannot be used interchangeably.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Perfusión
/
Volumen Sistólico
/
Ecocardiografía
/
Tomografía Computarizada de Emisión de Fotón Único
/
Estudios Retrospectivos
/
Ventrículos Cardíacos
Tipo de estudio:
Estudio observacional
Límite:
Femenino
/
Humanos
Idioma:
Coreano
Revista:
Nuclear Medicine and Molecular Imaging
Año:
2008
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS