Long-Term Stability of HBV, HCV, and HIV-1 National Reference Standards for in vitro Diagnostic Medical Devices Intended to Be Used for the Nucleic Acid Amplification Test / 대한수혈학회지
Korean Journal of Blood Transfusion
;
: 171-181, 2018.
Artículo
en Coreano
| WPRIM
| ID: wpr-716145
ABSTRACT
BACKGROUND:
National reference standards are essential to the quality assessment and regulatory approval of in vitro diagnostic medical devices. However, the long-term stability of national reference standards has not been comprehensively secured. This study was performed to assessment on the long-term stability of the hepatitis B virus (HBV), hepatitis C virus (HCV), and human immunodeficiency virus-1 (HIV-1) national reference standards intended to be used for the nucleic acid amplification test (NAT).METHODS:
The viral loads of the MFDS (Korea Ministry of Food and Drug Safety) working standard and recombinant DNA for HBV, HCV, and HIV-1 were measured before and after storage at −70℃ for up to 72 months using Cobas Ampliprep/Cobas Taqman assays (Roche Molecular System, Inc., Branchburg, USA) at defined time points.RESULTS:
The viral loads of national reference standards for in vitro diagnostic medical devices of HBV, HCV, and HIV-1 stored at −70℃ for up to 72 months did not differ significantly from the baseline viral load. The changes in viral load of national reference standards of HBV, HCV, and HIV-1 tested after storage at −70℃ for up to 72 months ranged from −0.36 to 0.16 log10 IU/mL and did not exceed 0.5 log10, which is the estimated intra-assay variation of molecular tests.CONCLUSION:
The HBV, HCV, and HIV-1 national reference standards for in vitro diagnostic medical devices intended to be used for the NAT were relatively stable after long-term storage at −70℃ for up to 72 months, regardless of the initial titer.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Juego de Reactivos para Diagnóstico
/
Técnicas In Vitro
/
ADN Recombinante
/
Virus de la Hepatitis B
/
VIH-1
/
Hepacivirus
/
Carga Viral
/
Técnicas de Amplificación de Ácido Nucleico
Tipo de estudio:
Estudio diagnóstico
/
Guía de Práctica Clínica
Límite:
Humanos
Idioma:
Coreano
Revista:
Korean Journal of Blood Transfusion
Año:
2018
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS