Your browser doesn't support javascript.
loading
Clinical Utility and Cross-Reactivity of Insulin and C-Peptide Assays by the Lumipulse G1200 System
Annals of Laboratory Medicine ; : 530-537, 2018.
Artículo en Inglés | WPRIM | ID: wpr-718330
ABSTRACT

BACKGROUND:

Measurement of insulin and C-peptide concentrations is important for deciding whether insulin treatment is required in diabetic patients. We aimed to investigate the analytical performance of insulin and C-peptide assays using the Lumipulse G1200 system (Fujirebio Inc., Tokyo, Japan).

METHODS:

We examined the precision, linearity, and cross-reactivity of insulin and C-peptide using five insulin analogues and purified proinsulin. A method comparison was conducted between the Lumipulse G1200 and Roche E170 (Roche Diagnostics, Mannheim, Germany) systems in 200 diabetic patients on insulin treatment. Reference intervals for insulin and C-peptide concentrations were determined in 279 healthy individuals.

RESULTS:

For insulin and C-peptide assays, within-laboratory precision (% CV) was 3.78–4.14 and 2.89–3.35%, respectively. The linearity of the insulin assay in the range of 0–2,778 pmol/L was R2=0.9997, and that of the C-peptide assay in the range of 0–10 nmol/L was R2=0.9996. The correlation coefficient (r) between the Roche E170 and Lumipulse G1200 results was 0.943 (P < 0.001) for insulin and 0.996 (P < 0.001) for C-peptide. The mean differences in insulin and C-peptide between Lumipulse G1200 and the Roche E170 were 19.4 pmol/L and 0.2 nmol/L, respectively. None of the insulin analogues or proinsulin showed significant cross-reactivity with the Lumipulse G1200. Reference intervals of insulin and C-peptide were 7.64–70.14 pmol/L and 0.17–0.85 nmol/L, respectively.

CONCLUSIONS:

Insulin and C-peptide tests on the Lumipulse G1200 show adequate analytical performance and are expected to be acceptable for use in clinical areas.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Proinsulina / Péptido C / Diabetes Mellitus / Insulina / Métodos Límite: Humanos Idioma: Inglés Revista: Annals of Laboratory Medicine Año: 2018 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Proinsulina / Péptido C / Diabetes Mellitus / Insulina / Métodos Límite: Humanos Idioma: Inglés Revista: Annals of Laboratory Medicine Año: 2018 Tipo del documento: Artículo