Your browser doesn't support javascript.
loading
Traumatic Disassembly of a Bipolar Hip Prosthesis: A Case Report
Journal of the Korean Hip Society ; : 151-154, 2011.
Artículo en Coreano | WPRIM | ID: wpr-727208
ABSTRACT
We report a rare case of a complete disassembly of the inner head from the bipolar cup without polyethylene wear and locking system failure. An 84-year-old man who had a femur neck fracture of the right hip underwent a replacement of the bipolar prosthesis 4 months ago. He fell down from a height of a chair and felt a sudden sharp pain in the right hip. Roentgenograms revealed that the inner head was dislocated from the outer cup; subsequently, a revision surgery was carried out. When the outer head was removed from the acetabulum, the bearing insert and locking ring were not deformed. Because there was no evidence of loosening of the femoral stem, identicallysized bipolar cup and metal head replacement was performed. At a 6 months follow-up, he had mild hip pain but had returned fully to daily living.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Prótesis e Implantes / Ursidae / Estudios de Seguimiento / Polietileno / Luxaciones Articulares / Fracturas del Cuello Femoral / Hemiartroplastia / Cabeza / Cadera / Acetábulo Tipo de estudio: Estudio observacional / Estudio pronóstico Límite: Anciano / Aged80 / Humanos Idioma: Coreano Revista: Journal of the Korean Hip Society Año: 2011 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Prótesis e Implantes / Ursidae / Estudios de Seguimiento / Polietileno / Luxaciones Articulares / Fracturas del Cuello Femoral / Hemiartroplastia / Cabeza / Cadera / Acetábulo Tipo de estudio: Estudio observacional / Estudio pronóstico Límite: Anciano / Aged80 / Humanos Idioma: Coreano Revista: Journal of the Korean Hip Society Año: 2011 Tipo del documento: Artículo