In Vitro Observation of Air Bubbles during Delivery of Various Detachable Aneurysm Embolization Coils
Korean Journal of Radiology
;
: 412-416, 2012.
Artículo
en Inglés
| WPRIM
| ID: wpr-72934
ABSTRACT
OBJECTIVE:
Device- or technique-related air embolism is a drawback of various neuro-endovascular procedures. Detachable aneurysm embolization coils can be sources of such air bubbles. We therefore assessed the formation of air bubbles during in vitro delivery of various detachable coils. MATERIALS ANDMETHODS:
A closed circuit simulating a typical endovascular coiling procedure was primed with saline solution degassed by a sonification device. Thirty commercially available detachable coils (7 Axium, 4 GDCs, 5 MicroPlex, 7 Target, and 7 Trufill coils) were tested by using the standard coil flushing and delivery techniques suggested by each manufacturer. The emergence of any air bubbles was monitored with a digital microscope and the images were captured to measure total volumes of air bubbles during coil insertion and detachment and after coil pusher removal.RESULTS:
Air bubbles were seen during insertion or removal of 23 of 30 coils (76.7%), with volumes ranging from 0 to 23.42 mm3 (median 0.16 mm3). Air bubbles were observed most frequently after removal of the coil pusher. Significantly larger amounts of air bubbles were observed in Target coils.CONCLUSION:
Variable volumes of air bubbles are observed while delivering detachable embolization coils, particularly after removal of the coil pusher and especially with Target coils.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Imagen por Resonancia Magnética
/
Estadísticas no Paramétricas
/
Medición de Riesgo
/
Embolia Intracraneal
/
Embolia Aérea
/
Embolización Terapéutica
/
Microscopía
Tipo de estudio:
Estudio de etiología
/
Factores de riesgo
Idioma:
Inglés
Revista:
Korean Journal of Radiology
Año:
2012
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS