ABSTRACT
Objective To explore the advantages and security of orthopaedic TiRobot used to assist internal fixation with sacroiliac lag screws.Methods From December 2015 to August 2017,11 patients with fracture of pelvic posterior ring or sacroiliac separation were treated by internal fixation with sacroiliac lag screws assisted by an orthopaedic TiRobot at Department of Orthopaedic Surgery,Affiliated Hospital to Chengdu University.They were 7 men and 4 women,aged from 23 to 61 years (average,42.5 years).According to AO classification,there were one case of type B1.2,one case of type B2.2,6 cases of type C1.2,2 cases of type C1.3 and one case of type C2.3.The number of sacroiliac lag screws inserted,time for planning insertion approaches,fluoroscopy frequency,fluoroscopy time,exposure time of C-arm X-ray,operation time,blood loss and reduction quality were documented.Results The 11 patients were followed up for 5 to 22 months (average,9.5 months).A total of 15 sacroiliac screws were inserted.In the operations,time for planning insertion approaches averaged 8.5 minutes (from 9 to 25 minutes),fluoroscopy frequency 8.5 times (from 4 to 15 times),fluoroscopy time 5.5 minutes (from 3.2 to 6.5 minutes),exposure time of C-arm X-ray machine 5.8 seconds (from 2.4 to 16.3 seconds),operation time 34.5 minutes (from 25 to 45 minutes),and blood loss 35 mL (from 10 to 80 mL).All the screw positions were satisfactory,with no penetration into the sacral canal,sacral foramen or bone cortex.No postoperative neurovascular injury happened.All the incisions primarily healed.All the fractures united well after 4 to 7 months (average,5.6 months),without any screw loosening.By the Matta scoring for fracture reduction,9 cases were excellent,one was good and one fair.By the Majeed functional scoring for pelvic fractures,8 cases were rated as excellent and 3 as good at the last follow-up.Conclusion Orthopaedic TiRobots can be used to assist internal fixation with sacroiliac lag screws in the treatment of pelvic posterior ring injuries,with advantages of limited invasion,time,hemorrhage,and high safety and accuracy as well.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Orthopaedic Trauma Año: 2019 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Orthopaedic Trauma Año: 2019 Tipo del documento: Artículo