Assessment of a portable monitoring device Watch PAT 200 in the diagnosis of obstructive sleep apnea hypopnea syndrome / 临床耳鼻咽喉头颈外科杂志
Journal of Clinical Otorhinolaryngology Head and Neck Surgery
;
(24): 1343-1347, 2013.
Artículo
en Chino
| WPRIM
| ID: wpr-747118
ABSTRACT
OBJECTIVE@#To assess the accuracy of a wrist worn Watch PAT 200 in diagnosing obstructive sleep apnea hypopnea syndrome (OSAHS) by comparing with the standard polysomnography (PSG).@*METHOD@#Twenty-eight adults with suspected OSAHS underwent a standard in-lab PSG while wearing a Watch PAT 200. PSG events were manually scored according to standard criteria (AASM). Watch PAT data were collected including changes of apnea hypopnea, sleep stages, peripheral arterial volume, signal oxygen saturation and heart rate, and then apnea hypopnea index (AHI) was analyzed by an automatic algorithm.@*RESULT@#For PSG, the mean score of AHI was (23.00 +/- 21.55)/h, and for Watch PAT, a mean score of AHI was (25.99 +/- 19.09)/h. There is a statistically positive correlation between PSG-AHI and PAT-AHI (r = 0.92, P < 0.01). The Coincidence rate of the sleep-wake assessment based on 30-second bins between the PSG and Watch PAT 200 was (89 +/- 6)%.@*CONCLUSION@#Watch PAT 200 could detected OSAHS based on AHI with comparable accuracy with standard PSG. And it would provide a reasonably accurate estimation of sleep and wakefulness stages in patients with OSAHS.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Fases del Sueño
/
Polisomnografía
/
Monitoreo Ambulatorio
/
Apnea Obstructiva del Sueño
/
Diagnóstico
Tipo de estudio:
Estudio diagnóstico
Límite:
Adulto
/
Anciano
/
Aged80
/
Femenino
/
Humanos
/
Masculino
Idioma:
Chino
Revista:
Journal of Clinical Otorhinolaryngology Head and Neck Surgery
Año:
2013
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS