Radiological Stability after Revision of Infected Total Knee Arthroplasty Using Modular Metal Augments
The Journal of Korean Knee Society
;
: 55-61, 2016.
Artículo
en Inglés
| WPRIM
| ID: wpr-759204
ABSTRACT
PURPOSE:
To evaluate the radiological stability according to the number of modular augments after revision of infected total knee arthroplasty (TKA). MATERIALS ANDMETHODS:
Between February 2006 and September 2013, 37 patients (39 knees) followed > or =2 years after revision of infected TKA using modular metal augments for bone defects were reviewed retrospectively. We divided the patients into 3 groups according to the number of augments into group A ( or = augments, 7 knees) and evaluated the width of radiolucent zones around the implant at the last follow-up.RESULTS:
There were 3 Anderson Orthopedic Research Institute type I, 33 type II, and 3 type III bone defects. The mean number of radiolucent zones of group A was 3 and the sum of width averaged 4.4 mm. In group B, the values were 4.8 and 6.2 mm, respectively. In group C, the values were 8.1 and 12.9 mm, respectively. The differences between the three groups were statistically significant.CONCLUSIONS:
In revision TKA with modular metal augmentation caused by infected TKA, increased modularity can result in radiological instability.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Ortopedia
/
Artroplastia
/
Estudios Retrospectivos
/
Estudios de Seguimiento
/
Academias e Institutos
/
Rodilla
Tipo de estudio:
Estudio observacional
/
Estudio pronóstico
Límite:
Humanos
Idioma:
Inglés
Revista:
The Journal of Korean Knee Society
Año:
2016
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS