OBJECTIVES:
The aim of this study was to test the applicability of haptic feedback using a smartwatch to the delivery of cardiac compression (CC) by professional healthcare providers.METHODS:
A prospective, randomized, controlled, case-crossover, standardized simulation study of 20 medical professionals was conducted. The participants were randomly assigned into haptic-first and non-haptic-first groups. The primary outcome was an adequate rate of 100–120/min of CC. The secondary outcome was a comparison of CC rate and adequate duration between the good and bad performance groups.RESULTS:
The mean interval between CCs and the number of haptic and non-haptic feedback-assisted CCs with an adequate duration were insignificant. In the subgroup analysis, both the good and bad performance groups showed a significant difference in the mean CC interval between the haptic and non-haptic feedback-assisted CC groups—good haptic feedback-assisted (0.57–0.06) vs. non-haptic feedback-assisted (0.54–0.03), p < 0.001; bad haptic feedback-assisted (0.57–0.07) vs. non-haptic feedback-assisted (0.58–0.18), p = 0.005—and the adequate chest compression number showed significant differences— good haptic feedback-assisted (1,597/75.1%) vs. non-haptic feedback-assisted (1,951/92.2%), p < 0.001; bad haptic feedbackassisted (1,341/63.5%) vs. non-haptic feedback-assisted (523/25.4%), p < 0.001.CONCLUSIONS:
A smartwatch cardiopulmonary resuscitation feedback system could not improve rescuers' CC rate. According to our subgroup analysis, participants might be aided by the device to increase the percentage of adequate compressions after one minute.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Tórax
/
Estudios Prospectivos
/
Reanimación Cardiopulmonar
/
Personal de Salud
/
Entrenamiento Simulado
/
Teléfono Inteligente
/
Masaje Cardíaco
Tipo de estudio:
Ensayo Clínico Controlado
/
Estudio observacional
Límite:
Humanos
Idioma:
Inglés
Revista:
Healthcare Informatics Research
Año:
2019
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS