Artificial Urinary Sphincter Cuff Size Predicts Outcome in Male Patients Treated for Stress Incontinence: Results of a Large Central European Multicenter Cohort Study / 대한배뇨장애요실금학회지
International Neurourology Journal
; : 219-225, 2019.
Article
en En
| WPRIM
| ID: wpr-764121
Biblioteca responsable:
WPRO
ABSTRACT
PURPOSE: The aim was to study the correlation between cuff size and outcome after implantation of an AMS 800 artificial urinary sphincter. METHODS: A total of 473 male patients with an AMS 800 sphincter implanted between 2012 and 2014 were analyzed in a retrospective multicenter cohort study performed as part of the Central European Debates on Male Incontinence (DOMINO) Project. RESULTS: Single cuffs were implanted in 54.5% and double cuffs in 45.5% of the patients. The cuffs used had a median circumference of 4.5 cm. Within a median follow of 18 months, urethral erosion occurred in 12.8% of the cases and was associated significantly more often with small cuff sizes (P<0.001). Multivariate analysis showed that, apart from cuff size (P=0.03), prior irradiation (P<0.001) and the penoscrotal approach (P=0.036) were associated with an increased erosion rate. Continence rate tended to be highest with median cuff sizes (4–5.5 cm). CONCLUSIONS: Apart from irradiation and the penoscrotal approach, small cuff size is a risk factor for urethral erosion. Results are best with cuff sizes of 4.5–5.5 cm.
Palabras clave
Texto completo:
1
Índice:
WPRIM
Asunto principal:
Análisis Multivariante
/
Estudios Retrospectivos
/
Factores de Riesgo
/
Estudios de Cohortes
/
Esfínter Urinario Artificial
Tipo de estudio:
Etiology_studies
/
Incidence_studies
/
Observational_studies
/
Prognostic_studies
Límite:
Humans
/
Male
Idioma:
En
Revista:
International Neurourology Journal
Año:
2019
Tipo del documento:
Article