Axisymmetric Contact Stress analysis of an Artificial Hip Joint of the Conical Fitting Type / 대한정형외과학회잡지
The Journal of the Korean Orthopaedic Association
;
: 1665-1672, 1989.
Artículo
en Coreano
| WPRIM
| ID: wpr-769123
ABSTRACT
The purpose of this study is to investigate the effect of the ball-cone contact interface variation on the ceramic ball stress which is the dominant factor of the design of the moldular artificial hip joint. A contact stress analysis problem is formulated using the optimization principle and finite element analysis technique on the ball-cone system of three dimensional axi-symmetrical model, Numerical procedure is employed to get the solution for varying angles of cones and angular mismatches between the ball and cone interfaces. There is a fair amount of friction in the interface and included in the analysis model.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Cerámica
/
Fricción
/
Análisis de Elementos Finitos
/
Cadera
/
Articulación de la Cadera
Tipo de estudio:
Estudio pronóstico
Idioma:
Coreano
Revista:
The Journal of the Korean Orthopaedic Association
Año:
1989
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS