ABSTRACT
The quality inspection of magnetic resonance imaging (MRI) performance parameters is an important means to ensure the image quality and the reliability of diagnosis results. There are some problems in the manual calculation and eye recognition of the quality inspection parameters, such as strong subjectivity and low efficiency. In view of these facts, an automatic analysis system for MRI quality detection based on QT is proposed and implemented in C++ language. The image processing algorithm is introduced to automatically measure and calculate the quality inspection parameters. The software with comprehensive functions is designed to systematically manage the quality inspection information of MRI. The experimental results show that the automatically calculated parameters are consistent with the manually calculated ones. Accordingly, the accuracy and reliability of the algorithm is verified. The whole system is efficient, convenient and easy to operate, and it can meet the actual needs of MRI quality inspection.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Estándares de Referencia
/
Algoritmos
/
Procesamiento de Imagen Asistido por Computador
/
Imagen por Resonancia Magnética
/
Reproducibilidad de los Resultados
Tipo de estudio:
Estudio diagnóstico
/
Guía de Práctica Clínica
Idioma:
Chino
Revista:
Journal of Biomedical Engineering
Año:
2019
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS