Device and method for evaluating the performance of dry electrode for fabric surface bioelectricity / 生物医学工程学杂志
J. biomed. eng
; Sheng wu yi xue gong cheng xue za zhi;(6): 460-467, 2019.
Article
en Zh
| WPRIM
| ID: wpr-774184
Biblioteca responsable:
WPRO
ABSTRACT
In order to quantitatively evaluate the performance of dry electrode for fabric surface bioelectricity, a set of active measuring devices that can simulate electrocardiosignal has been developed on the basis of passive system by our group. Five Ag/AgCl fabric dry electrodes were selected to test and evaluate the devices. The results show that the deviation ratios of peak time interval of the five electrodes are all less than 1%. The maximum voltage amplitude decay rate is 7.2%, and the noise amplitudes are lower than 0.004 mV. The variable coefficient of peak time offset is less than 8%. The variable coefficient of voltage amplitude is less than 2%. The variable coefficient of noise amplitude is less than 10%. Research shows the devices has good repeatability and stability in measuring the simulated electrocardiosignal. The active measuring devices proposed in this paper can provide a new method for performance evaluation and standard formulation of surface bioelectricity dry electrode.
Palabras clave
Texto completo:
1
Índice:
WPRIM
Asunto principal:
Textiles
/
Electrodos
/
Dispositivos Electrónicos Vestibles
Idioma:
Zh
Revista:
J. biomed. eng
/
Sheng wu yi xue gong cheng xue za zhi
Año:
2019
Tipo del documento:
Article