Acceptance Test and Image Quality Assurance of MRI Simulator Equipment / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 455-459, 2018.
Artículo
en Chino
| WPRIM
| ID: wpr-775529
ABSTRACT
MRI simulator(MRI-Sim) images have unique clinical advantages with higher resolution of soft tissue and clearer visualization of tissue boundaries. Thus, the precise positioning of the tumor target area can be achieved and it is widely used in the field of radiotherapy. This article focuses on the acceptance test project and image quality assurance work of MRI-Sim equipment. The obtained ACR phantom images were used to analyze various image quality assurance indicators, and the results all reached the set standards, thereby ensuring that the obtained images meet the requirements of clinical applications.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Garantía de la Calidad de Atención de Salud
/
Imagen por Resonancia Magnética
/
Fantasmas de Imagen
Tipo de estudio:
Guía de Práctica Clínica
Idioma:
Chino
Revista:
Chinese Journal of Medical Instrumentation
Año:
2018
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS