Your browser doesn't support javascript.
loading
Relationship between the Retinal Thickness Analyzer and the GDx VCC Scanning Laser Polarimeter, Stratus OCT Optical Coherence Tomograph, and Heidelberg Retina Tomograph II Confocal Scanning Laser Ophthalmoscopy
Korean Journal of Ophthalmology ; : 10-17, 2008.
Artículo en Inglés | WPRIM | ID: wpr-78062
ABSTRACT

PURPOSE:

To assess the relationship between the retinal thickness analyzer (RTA) parameters, and those of the GDx VCC scanning laser polarimeter (GDx VCC), Stratus OCT optical coherence tomography (Stratus OCT), and Heidelberg retinal tomograph II confocal scanning laser ophthalmoscopy (HRT II).

METHODS:

Twenty-nine primary open-angle glaucoma patients were retrospectively included in this study. Measurements were obtained using the RTA, GDx VCC, Stratus OCT, and HRT II. We calculated the correlation coefficients between the parameters of RTA and those of the other studies.

RESULTS:

Among the optic disc parameters of RTA, the cup volume was best correlated with Stratus OCT (R=0.780, p<0.001) and HRT II (R=0.896, p<0.001). Among the posterior pole retinal thickness parameters, the posterior pole abnormally thin area (PPAT) of the RTA and the inferior average of the GDx VCC were best correlated (R=-0.596, p=0.001). The PPAT of the RTA and the inferior maximum of the Stratus OCT were best correlated (R=-0.489, p=0.006). The perifoveal minimum thickness (PFMT) of the RTA and the cup shape measurement of the HRT II were best correlated (R=-0.565, p=0.004).

CONCLUSIONS:

Many RTA optic disc parameters were significantly correlated with those of the Stratus OCT and HRT II. The RTA posterior pole retinal thickness parameters were significantly correlated with those of the GDx VCC, Stratus OCT and HRT II. The RTA optic disc and posterior pole retinal thickness parameters may be valuable in the diagnosis of glaucoma.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Oftalmoscopía / Disco Óptico / Retina / Enfermedades del Nervio Óptico / Glaucoma de Ángulo Abierto / Estudios Retrospectivos / Tomografía de Coherencia Óptica / Técnicas de Diagnóstico Oftalmológico / Rayos Láser Tipo de estudio: Estudio observacional Límite: Adolescente / Adulto / Anciano / Femenino / Humanos / Masculino Idioma: Inglés Revista: Korean Journal of Ophthalmology Año: 2008 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Oftalmoscopía / Disco Óptico / Retina / Enfermedades del Nervio Óptico / Glaucoma de Ángulo Abierto / Estudios Retrospectivos / Tomografía de Coherencia Óptica / Técnicas de Diagnóstico Oftalmológico / Rayos Láser Tipo de estudio: Estudio observacional Límite: Adolescente / Adulto / Anciano / Femenino / Humanos / Masculino Idioma: Inglés Revista: Korean Journal of Ophthalmology Año: 2008 Tipo del documento: Artículo