Versa HD Volume-modulated Accelerator High Voltage Circuit Fault Maintenance / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 371-373, 2020.
Artículo
en Chino
| WPRIM
| ID: wpr-828184
ABSTRACT
This paper introduces a fast repair methods of Versa HD volume-modulated accelerator's high voltage circuit faultthe key points test method. To identify five key points①Enter maintenance mode to check for AFC deviation; ② The magnetron waveform MI and PFN waveform PFN V are detected on the maintenance terminal board; ③ Detect thyratron waveform; ④ Check the supply voltage of thyratron; ⑤ HT PSU 600 V DC test 600 V normal or not. The waveform or voltage is also measured to efficiently narrow the fault range to find out the cause of the fault, quickly remove the fault.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Mantenimiento
Idioma:
Chino
Revista:
Chinese Journal of Medical Instrumentation
Año:
2020
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS