Your browser doesn't support javascript.
loading
Comparison of Body Fluid Differential CountsUsing a Manual Counting Method oran Automated Hematology Analyzer
Journal of Laboratory Medicine and Quality Assurance ; : 26-32, 2020.
Artículo | WPRIM | ID: wpr-836059
ABSTRACT
Background@#Two methods of counting cells in body fluids were compared;manual counting using a Neubauer chamber, and automated cell countingusing an XN-350 hematology analyzer. @*Methods@#Cells from 32 body fluid samples were counted by manualexamination and by an automated analyzer. Total cells (TC), white bloodcells (WBC), red blood cells (RBC), polymorphonuclear leukocytes (PMN),mononuclear leukocytes (MN), neutrophils, lymphocytes, monocytes, andeosinophils were each counted by both methods. The results were comparedusing the Pearson correlation test, Bland-Altman regression analysis, andPassing-Bablok regression analysis. @*Results@#The two methods showed very strong correlation in TC, WBC,RBC, PMN, and MN counts, strong correlation in % neutrophils, and %lymphocytes, and weak correlation in % monocytes and % eosinophils.Using Bland-Altman regression analysis, the mean biases for TC, WBC, andRBC were -270, -257.4, and -1,256.09, respectively, and 0.15 for PMN andMN. Research parameters were compared as well mean biases were -1.31,-2.46, -5.16, and -3.58 for % neutrophils, % monocytes, % lymphocytes,and % eosinophils, respectively. Passing-Bablok regression equationswere y=1.039x+20, y=1.037x+19, y=1.259x+0.0, y=0.983x+1.541, andy=0.983x+0.125 for TC, WBC, RBC, PMN, and MN, respectively. The equationswere y=0.955x+2.194 for % neutrophils, y=0.965x+1.184 for % monocytes,y=1.003x+0.161 for % lymphocytes, and y=x+0.75 for % eosinophils. @*Conclusions@#WBC differential count results performed by an automatedhematology analyzer generally show good correlation with our referencemethod, Neubauer chamber counting.
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Revista: Journal of Laboratory Medicine and Quality Assurance Año: 2020 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Revista: Journal of Laboratory Medicine and Quality Assurance Año: 2020 Tipo del documento: Artículo