Your browser doesn't support javascript.
loading
Acceptance tests of beam performance for pencil beam scanning (PBS) proton system of IBA / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology ; (6): 1075-1079, 2020.
Artículo en Chino | WPRIM | ID: wpr-868743
ABSTRACT

Objective:

To explore the beam performance of GTR 360° PBS dedicated nozzle for IBA Proteus Plus proton system and evaluate the beam characteristics of PBS during acceptance test procedure (ATP) to meet the precision and safety requirements of clinical treatment.

Methods:

According to the requirements of acceptance tests, the beam tests of PBS mainly included test of integrated depth dose (IDD), test of maximum field size at the maximum and minimum ranges, field uniformity and informative PBS tests for the GTR 360°, test of lateral dose uniformity in a single field uniform dose in a cube, test of longitudinal dose uniformity in a single field uniform dose in a cube, and test of monitor unit reproducibility and proportionality.

Results:

The maximum deviation measured of range, fall back accuracy and distal fall-off was 0.03 g/m 2, 0.01 g/m 2 and 0.078 g/m 2, respectively. Maximum field size at the maximum and minimum ranges was 30.2 cm×40.2 cm and 30.1 cm×40.1 cm, respectively. During the different gantry angles and ranges, the lowest γ passing rate of a pattern plan was 97%, the maximum deviation of the centrical beam spot size at the X and Y axes was -0.16 mm and -0.21 mm, and the worst symmetry was 0.80%. Compared with the centrical beam spot, the maximum size deviation of the other beam spots was 0.11 mm and 0.14 mm at the X and Y axes, and the maximum position accuracy deviation of the beam spot was 0.60 mm and 0.43 mm at the X and Y axes. The maximum deviation of lateral dose uniformity at at the X and Y axes was 0.55% and 0.80% in the high energy region, and 0.6% and 0.75% in the low energy region. The maximum deviation of longitudinal dose uniformity was 0.79% in the high energy region, and 2.22% in the low energy region. The monitor unit (MU) reproducibility factor was 0.106% and the maximum proportionality deviation was 0.67%.

Conclusion:

The dedicated nozzle of PBS has passed all the beam performance acceptance tests, which meet the requirements of all parameters, and the whole system yields relatively high accuracy, repeatability and good stability.
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiation Oncology Año: 2020 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiation Oncology Año: 2020 Tipo del documento: Artículo