LC-MS/MS method for determination of salidroside in the capsule / 药学实践杂志
Article en Zh | WPRIM | ID: wpr-875678
Biblioteca responsable: WPRO
ABSTRACT
Objective To establish a LC-MS/MS method for the determination of salidroside in the capsule. Methods An electrospray ionization and multiple reaction detection were used to detect negative ion. Theophylline was used as standard. The detection ions of salidroside and theophylline used for quantitative analysis were m/z 299.0→119.0, and m/z 178.8→164.0, respectively. The Shim-pack XR-ODS (3.0 mm×75 mm, 2.0 μm) column was used for separation. The mobile phase was acetonitrile: 5 mmol/L ammonium acetate solution (90∶10, V/V). The flow rate was 0.40 ml/min. The column temperature was 25 ℃. Results The content of salidroside was analyzed within 3 minutes. The linear range was 10–2 000 ng/ml, and the minimum detection limit was 10 ng/ml. Conclusion The method has good repeatability, high sensitivity, fast analysis speed and simple operation. It can be used as a method for the determination of salidroside in the capsule. It is suitable for the quality inspection of drugs and convenient for safe use.