Simultaneous detection of 6 isomers of dimethylaniline in workplace air by gas chromatography / 中国职业医学
China Occupational Medicine ; (6): 572-577, 2020.
Artículo en Chino | WPRIM | ID: wpr-881940
ABSTRACT

OBJECTIVE:

To establish a method for simultaneous detection of 6 isomers of dimethylaniline(DMA) in workplace air by gas chromatography(GC).

METHODS:

The vapor and aerosol DMA in workplace air were collected by GDH-1 air sampling tube, and desorbed and eluted with absolute ethanol, then separated by a special capillary column for amines analysis, and finally detected by GC-flame ionization detector.

RESULTS:

The quantitative detection range of 6 isomers of DMA was 0.26-226.66 mg/L, with all the correlation coefficients greater than 0.999 00. The minimum detectable concentration was 0.01-0.02 mg/m~3, and the minimum quantification concentration was 0.04-0.05 mg/m~(3 )(15.00 L sample, 2.00 mL sample solution). The average desorption and elution efficiency was 92.48%-104.60%. The within-run relative standard deviations(RSD) was 0.31%-4.51%, and the between-run RSD was 1.23%-6.03%. The samples can be sealed and stored for more than 19 days at room temperature.

CONCLUSION:

The method is suitable for simultaneous detection of 6 isomers of DMA in workplace air.