Your browser doesn't support javascript.
loading
Comparison of Icare Rebound Tonometer with Goldmann Applanation Tonometry
Journal of the Korean Ophthalmological Society ; : 296-302, 2013.
Artículo en Coreano | WPRIM | ID: wpr-88444
ABSTRACT

PURPOSE:

To compare Icare rebound tonometer (IRT) and Goldmann applanation tonometer (GAT) and investigate the clinical usefulness of IRT.

METHODS:

In a retrospective study of 131 eyes with glaucoma and glaucoma suspect, intraocular pressure (IOP) was measured with IRT and GAT. The correlation between IRT and GAT and the influence of refractive error (spherical equivalent; SE) and central corneal thickness (CCT) were analyzed.

RESULTS:

A distinct correlation between IRT and GAT was found. IOP measured with GAT was 2.23 +/- 3.16 mm Hg higher than with IRT (p or = -5 diopter) was measured relatively lower than in hyperopic eyes (SE < -5 diopter) with GAT (+1.50 +/- 0.68 mm Hg) and IRT (+1.88 +/- 0.75 mm Hg).

CONCLUSIONS:

IRT showed strong correlation with GAT, but IOP measured approximately 2 mm Hg lower. The IOP tends to measure lower at low CCT and in myopic eyes.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Péptidos / Errores de Refracción / Glaucoma / Estudios Retrospectivos / Ojo / Presión Intraocular / Complejo Hierro-Dextran / Manometría Tipo de estudio: Estudio observacional Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Año: 2013 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Péptidos / Errores de Refracción / Glaucoma / Estudios Retrospectivos / Ojo / Presión Intraocular / Complejo Hierro-Dextran / Manometría Tipo de estudio: Estudio observacional Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Año: 2013 Tipo del documento: Artículo