Your browser doesn't support javascript.
loading
Helical Blade Locking Sleeve Disassembly Following Failed Femur Intertrochanter Fracture
Journal of the Korean Fracture Society ; : 112-116, 2021.
Artículo en Inglés | WPRIM | ID: wpr-900796
ABSTRACT
A helical blade type of hip screw is used widely for the operative management of femoral trochanteric fractures. A 73-year-old female patient was admitted for femoral trochanteric fracture AO/OTA type 31A2.2. A helical blade locking sleeve dissembled 18 weeks after surgery did not achieve accurate reduction. The patient underwent bipolar hip hemiarthroplasty because the fracture reduction was lost, and it was impossible to remove the remaining helical blade without bone loss. The authors report this case of a rare complication of helical blade.
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Inglés Revista: Journal of the Korean Fracture Society Año: 2021 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Inglés Revista: Journal of the Korean Fracture Society Año: 2021 Tipo del documento: Artículo