ABSTRACT
There have been some reports of gross trunnion failure (GTF) in total hip arthroplasty. Here, we report a case of GTF 19 years after bipolar hemiarthroplasty using a 28-mm head with a 14/16 taper bore. Compared to other GTF reports, the current case had some unusual aspects bipolar hemiarthroplasty, 28-mm head, relatively lateonset, and no apparent findings of metallosis though a severe one was evident. A Computed tomography scout view provided valuable information in evaluating polyethylene, metal head, and neck inside the bipolar cup. The current report suggests a need for concerned regarding short trunnion length which may be associated with GTF.
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Inglés Revista: Hip & Pelvis Año: 2021 Tipo del documento: Artículo