Structural Design and Finite Element Analysis of New Type Posterior Integral Glaucoma Valve / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
;
(6): 292-295, 2022.
Artículo
en Chino
| WPRIM
| ID: wpr-928907
ABSTRACT
The treatment of refractory Glaucoma is a difficult problem in clinical ophthalmology. For refractory glaucoma patients with hyphema, shallow anterior chamber, anterior conglutination of peripheral chamber angle, corneal endothelium dystrophy or decompensated, at present, there is no effective treatment. In order to solve this problem, a new type posterior integral glaucoma valve with IOP control device was designed using medical titanium alloy, and the valve model was established by Abaqus software, and the stiffness and preload of the valve were analyzed by finite element method. The results showed that the opening and closing of the valve were controlled automatically by the pressure difference between the front and back of the valve, and the opening and flow rate of the valve increase dynamically with the increase of intraocular pressure, and finally reached the set ideal IOP value of steady state.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Glaucoma
/
Estudios de Seguimiento
/
Resultado del Tratamiento
/
Implantes de Drenaje de Glaucoma
/
Análisis de Elementos Finitos
/
Presión Intraocular
Tipo de estudio:
Estudio observacional
/
Estudio pronóstico
Límite:
Humanos
Idioma:
Chino
Revista:
Chinese Journal of Medical Instrumentation
Año:
2022
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS