Your browser doesn't support javascript.
loading
Accuracy of photogrammetry and conventional impression techniques for complete-arch implant rehabilitation: an in vitro comparative study / 中华口腔医学杂志
Chinese Journal of Stomatology ; (12): 168-172, 2022.
Artículo en Chino | WPRIM | ID: wpr-935844
ABSTRACT

Objective:

To compare the accuracy of photogrammetry and conventional impression techniques for complete-arch implant rehabilitation.

Methods:

An edentulous maxillary stone cast containing 8 screw-retained implant abutment replicas was derived from a 74-year-old male patient who visited the Department of Dental Implant Center, Capital Medical University School of Stomatology in September 2019. The stone cast was copied through the open-tray splinted impression, and the copied cast was used as the master cast for this study. The abutment-level impressions of master cast were made by photogrammetry (PG) and the conventional impression technique (CNV) by one attending doctor. Group PG after which scan bodies were connected to each implant replica, a photogrammetry system was used to obtain digital impressions of the master cast (n=10); Group CNV conventional open-tray splinted impression technique was performed to fabricate conventional definitive casts (n=10). After connecting the scan bodies onto each implant replicas, the master cast and the 10 definitive casts from group CNV were digitized with a laboratory reference scanner. All data of group PG, group CNV and mater cast were saved as ".stl" files. For all test scans and reference scan, the three-dimensional information of scan bodies were converted to implant abutment replicas using a computer aided design software (Exocad). The data of the group PG and the group CNV were respectively registered with the reference data (trueness analysis) and pairwise within group (precision analysis) for accuracy evaluation in a three-dimensional analysis software (Geomagic Control X).

Results:

The trueness and precision of group PG [(17.33±0.34) and (2.50±0.79) μm ] were significantly statistically better than those of group CNV [(24.30±4.16) and (26.12±4.54) μm] respectively (t=-5.29 and -34.35, P<0.001).

Conclusions:

For complete-arch implant abutment-level impression, photogrammetry produces significantly better accuracy than conventional impression technique.
Asunto(s)
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Fotogrametría / Implantes Dentales / Técnica de Impresión Dental / Diseño Asistido por Computadora / Materiales de Impresión Dental / Modelos Dentales Límite: Anciano / Humanos Idioma: Chino Revista: Chinese Journal of Stomatology Año: 2022 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Fotogrametría / Implantes Dentales / Técnica de Impresión Dental / Diseño Asistido por Computadora / Materiales de Impresión Dental / Modelos Dentales Límite: Anciano / Humanos Idioma: Chino Revista: Chinese Journal of Stomatology Año: 2022 Tipo del documento: Artículo