Hemolytic Anemia Case Caused by an Inverted Inner Felt after Bentall Operation
Journal of Korean Medical Science
; : 1827-1829, 2013.
Article
en En
| WPRIM
| ID: wpr-9505
Biblioteca responsable:
WPRO
ABSTRACT
A 26-yr-old male patient reported worsened dyspnea, dizziness one year after an emergency Bentall operation for type A aortic dissection. There was evidence of hemolytic anemia and aortogram revealed a significant stenosis at the distal anastomosis site. During the reoperation, we found the inner felt at the distal anastomosis was inverted causing a significant stenosis. The reoperation successfully resolved this problem. Here, we report a rare case of hemolytic anemia caused by an inverted inner felt after Bentall operation.
Palabras clave
Texto completo:
1
Índice:
WPRIM
Asunto principal:
Aneurisma de la Aorta
/
Complicaciones Posoperatorias
/
Reoperación
/
Factores de Tiempo
/
Prótesis Vascular
/
Anastomosis Quirúrgica
/
Ecocardiografía
/
Tomografía Computarizada por Rayos X
/
Enfermedad Aguda
/
Resultado del Tratamiento
Límite:
Adult
/
Humans
/
Male
Idioma:
En
Revista:
Journal of Korean Medical Science
Año:
2013
Tipo del documento:
Article