A case of recurrent lead fracture and complete dislocation after permanent pacemaker implantation / 대한내과학회지
Korean Journal of Medicine ; : 747-750, 2010.
Artículo en Coreano | WPRIM | ID: wpr-95595
ABSTRACT
With the increased use of implantable cardiac devices, the incidence of hardware problems has also increased. Some of the hardware problems might be induced by patient factors. We experienced recurrent pacemaker lead fracture and dislocation after permanent pacemaker implantation. The patient was a bus driver who used his left arm vigorously when he turned the steering wheel. After a new lead was inserted via an axillary vein approach and the patient changed his occupation, no more lead problems have occurred. When a recurrent hardware problem with an implantable pacemaker or defibrillator occurs, patient factors related to anatomy or behavior must be considered.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Brazo / Vena Axilar / Incidencia / Desfibriladores / Luxaciones Articulares / Bloqueo Atrioventricular / Ocupaciones Tipo de estudio: Estudio de incidencia / Estudio pronóstico Límite: Humanos Idioma: Coreano Revista: Korean Journal of Medicine Año: 2010 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Brazo / Vena Axilar / Incidencia / Desfibriladores / Luxaciones Articulares / Bloqueo Atrioventricular / Ocupaciones Tipo de estudio: Estudio de incidencia / Estudio pronóstico Límite: Humanos Idioma: Coreano Revista: Korean Journal of Medicine Año: 2010 Tipo del documento: Artículo