Your browser doesn't support javascript.
loading
Analysis of quality control and protection level test results of medical electron linear accelerators in Shanghai, China / 中国辐射卫生
Chinese Journal of Radiological Health ; (6): 10-14, 2023.
Artículo en Chino | WPRIM | ID: wpr-965364
ABSTRACT
@#<b>Objective</b> To investigate the quality control and protection level of medical electron linear accelerators in Shanghai, China. <b>Methods</b> The startified random sampling method was used to cover tertiary, secondary and ungraded hospitals, and a total of 30 medical electron linear accelerators in 15 hospitals were tested for quality control and protection level according to relevant standards. <b>Results</b> Five medical electron linear accelerators failed the quality control test, with an overall inspection pass rate of 83.3% and a re-inspection pass rate of 100%. The pass rate of flatness of square X-ray irradiation field (5 cm × 5 cm)-(30 cm × 30 cm) was 83.3%, the pass rate of symmetry of square X-ray irradiation field was 96.7%, and other indices were qualified. All medical electron linear accelerator rooms passed the protection test. <b>Conclusion</b> The protection of medical electron linear accelerator rooms in Shanghai meets the requirements of national standards, and some indices do not meet the requirements of national standards in the preliminary inspection. The quality control of medical electron linear accelerators should be further strengthened to ensure the treatment effect of patients.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Health Año: 2023 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Health Año: 2023 Tipo del documento: Artículo