Feasibility analysis of home-made AQA phantom applied for Cyberknife quality control / 中国辐射卫生
Chinese Journal of Radiological Health ; (6): 645-650, 2022.
Artículo en Chino | WPRIM | ID: wpr-965463
ABSTRACT
Objective@#To discuss the feasibility of applying home-made auto quality assurance (AQA) phantom in Cyberknife AQA test.@*Methods@#The same treatment plan for AQA test was designed in the Cyberknife treatment planning system. The AQA tests were performed under the same conditions using the home-made AQA phantom and the AQA phantom in the fourth-generation (G4) Cyberknife mode and the fifth-generation (VSI) Cyberknife mode, respectively, and each measurement was repeated five times to analyze and compare the differences in test results between the two phantoms.@*Results@#The total deviations of the home-made AQA phantom and the AQA phantom in the AQA test under the G4 Cy- berknife mode were (0.28 ± 0.12) mm and (0.28 ± 0.15) mm, respectively (P > 0.05); the test results under the VSI Cy- berknife mode were (0.46 ± 0.19) mm and (0.50 ± 0.07) mm, respectively (P > 0.05); the total deviations of all test results were smaller than 1.0 mm, which met the requirements of WS 667—2019 Specification for Testing of Quality Control in Ro- botic Arm Radiotherapy Device.@*Conclusion@#The test results of the home-made AQA phantom and the AQA phantom are consistent in the AQA test, and the home-made AQA phantom can be used for Cyberknife quality control.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Health Año: 2022 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Health Año: 2022 Tipo del documento: Artículo