Your browser doesn't support javascript.
loading
Influence of disinfection methods used in hospitals on monitoring personal dose of external irradiation during COVID-19 / 中国辐射卫生
Chinese Journal of Radiological Health ; (6): 98-101, 2023.
Artículo en Chino | WPRIM | ID: wpr-973159
ABSTRACT
@#<b>Objective</b> To study whether the disinfection methods for thermoluminescence dosimeters used by major hospitals during the COVID-19 pandemic had an impact on monitoring personal dose of external irradiation, and to screen for the disinfection procedures suitable for use during the pandemic. <b>Methods</b> Hospitals with fever clinics during the epidemic in Shandong Province were selected from March to October 2020, we investigated the disinfection methods as well as the types and frequency of use of disinfection supplies used for thermoluminescence dosimeters during the pandemic in hospitals. Simulations were performed in experimental and control groups. <b>Results</b> The average doses measured by the disc dosimeter with different disinfection methods and the control group were 0.92-0.99 mSv and 0.98 mSv, respectively, and the deviation was ≤ 6%. The average doses measured by the glass tube dosimeter with different disinfection methods and the control group were 0.20-0.22 and 0.21, respectively, and the deviation was ≤ 5%. <b>Conclusion</b> The results showed that the commonly used disinfection methods had no influence on personal dose monitoring, but some disinfection measures were not suitable for dosimeters.

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Health Año: 2023 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Chino Revista: Chinese Journal of Radiological Health Año: 2023 Tipo del documento: Artículo