ABSTRACT
@#<b>Objective</b> To analyze the process of intercomparison of national personal dose monitoring, evaluate the ability of personal dose monitoring, and ensure the accuracy and reliability of monitoring results in our laboratory. <b>Methods</b> In accordance with the intercomparison protocol for 2019—2021, an energy-discriminant thermoluminescence dosimeter was used for measurement at different doses. The uncertainty of measurement was evaluated and compared with the reference value. <b>Results</b> <i>H</i><sub>p</sub>(10) was measured for intercomparison in 2019—2021. In 2019, the single group performance difference was −0.02 to 0.02 and the comprehensive performance was 0.02. These values were 0.02-0.10 and 0.05 in 2020, and −0.02 to 0.02 and 0.01 in 2021. The intercomparison results were rated as excellent in the three consecutive years. <b>Conclusion</b> The personal dose monitoring system in our laboratory was in good condition, and the monitoring results were accurate and reliable. Improving the knowledge of personnel and cultivating a serious working attitude are important for intercomparison and personal dose monitoring.