Your browser doesn't support javascript.
loading
Efficacy of Biometry Using Swept-source Optical Coherence Tomography for Posterior Chamber Phakic Intraocular Lens Implantation
Journal of the Korean Ophthalmological Society ; : 374-381, 2023.
Artículo en Coreano | WPRIM | ID: wpr-977101
ABSTRACT
Purpose@#The proper implantable collamer lens (ICL) size affects ICL stability. This study compared device efficacy using white-to-white diameter (WTW) measurements with Orbscan II and IOL Master 700. @*Methods@#We retrospectively studied 90 eyes (45 patients) who underwent toric ICL implantation from January 2019 to February 2020 and were followed for 1 year. The correlation between WTW and anterior chamber depth (ACD) for each measuring device was analyzed. @*Results@#The mean WTW measured by IOL Master 700 and Orbscan II was 12.2 ± 0.3 and 11.6 ± 0.3 mm, respectively, while the mean ACD was 3.28 ± 0.16 and 3.20 ± 0.15 mm. The WTW and ACD measured with IOL Master 700 averaged 0.57 ± 0.12 and 0.08 ± 0.04 mm larger than with Orbscan II. The differences were significant and the regression analysis had high correlations (R2 = 0.875 and R2 = 0.913, respectively; both p < 0.001). @*Conclusions@#WTW measured by the IOL Master 700 can be used as a reference either alone or together with the Orbscan II value to determine ICL size.
Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Año: 2023 Tipo del documento: Artículo