The aortic connector system may reduce stroke during proximal venous anastomosis. However, the overall anastomotic patency rate has been generally reported to be low. From October 2002 to March 2004, 68 patients who received proximal anastomosis using the St. Jude Aortic Connector System were included in the study. There were 47 men and 21 women and their mean age was 65.68+/-6.68 yr old (52 to 85 yr). Grafts were evaluated by coronary angiography or multi-slice 16 channel 3-D CT at 6 days and at 6 months postoperatively. In the immediate postoperative period, no stenosis was observed by either angiography (n=22) or 3D CT (n=46). At 6-month postoperatively, we performed either angiography (n=7) or 3-D CT (n=52). Of these patients, 5 patients showed graft stenosis in the midportion, and 3 in the ostium. There were no stroke. Simple, and effective proximal anastomosis with good protection from cerebrovascular accident was achieved especially when calcification or atheromatous plaque was observed at the ascending aorta in the operation room. However, our mid term patency results raise concerns related to venous graft stenosis in the midportion. Therefore, longer follow up is recommended.
Texto completo:
Disponible
Índice:
WPRIM (Pacífico Occidental)
Asunto principal:
Aorta
/
Vena Safena
/
Anastomosis Quirúrgica
/
Puente de Arteria Coronaria
/
Estudios de Seguimiento
Tipo de estudio:
Estudio observacional
/
Estudio pronóstico
Límite:
Anciano
/
Aged80
/
Femenino
/
Humanos
/
Masculino
Idioma:
Inglés
Revista:
Journal of Korean Medical Science
Año:
2006
Tipo del documento:
Artículo
Similares
MEDLINE
...
LILACS
LIS