Your browser doesn't support javascript.
loading
Simulation Results and Analysis of Electric Field Distribution in Myocardial Tissue under Circular Electrode Electric Pulse Ablation / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 242-246, 2023.
Artículo en Chino | WPRIM | ID: wpr-982221
ABSTRACT
As a new energy source for atrial fibrillation ablation, electric pulse ablation has higher tissue selectivity and biosafety, so it has a great application prospect. At present, there is very limited research on multi-electrode simulated ablation of histological electrical pulse. In this study, a circular multi-electrode ablation model of pulmonary vein will be built on COMSOL5.5 platform for simulation research. The results show that when the voltage amplitude reaches about 900 V, it can make some positions achieve transmural ablation, and the depth of continuous ablation area formed can reach 3 mm when the voltage amplitude reaches 1 200 V. When the distance between catheter electrode and myocardial tissue is increased to 2 mm, a voltage of at least 2 000 V is required to make the depth of continuous ablation area reach 3 mm. Through the simulation of electric pulse ablation with ring electrode, the research results of this project can provide reference for the voltage selection in the clinical application of electric pulse ablation.
Asunto(s)

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Fibrilación Atrial / Ablación por Catéter / Electricidad / Electrodos / Frecuencia Cardíaca Límite: Humanos Idioma: Chino Revista: Chinese Journal of Medical Instrumentation Año: 2023 Tipo del documento: Artículo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponible Índice: WPRIM (Pacífico Occidental) Asunto principal: Fibrilación Atrial / Ablación por Catéter / Electricidad / Electrodos / Frecuencia Cardíaca Límite: Humanos Idioma: Chino Revista: Chinese Journal of Medical Instrumentation Año: 2023 Tipo del documento: Artículo