Your browser doesn't support javascript.
loading
Marcapasos bicamerales: resultados inmediatos y alejados / Short and long terms results of dual chamber pacemaker use
Dubernet Mazerolles, Juan; Fajuri Noemi, Alejandro; Acevedo B., Mónica; González M., Jorge A; Maturana Barahona, Gustavo; Zalaquett Sepúlveda, Ricardo; Garrido O., Luis; De la Paz P., Francisco; Casanegra Prjnat, Pablo.
  • Dubernet Mazerolles, Juan; Pontificia Universidad Católica de Chile. Hospital Clínico. Departamento de Enfermedades Cardiovasculares.
  • Fajuri Noemi, Alejandro; Pontificia Universidad Católica de Chile. Hospital Clínico. Departamento de Enfermedades Cardiovasculares.
  • Acevedo B., Mónica; Pontificia Universidad Católica de Chile. Hospital Clínico. Departamento de Enfermedades Cardiovasculares.
  • González M., Jorge A; Pontificia Universidad Católica de Chile. Hospital Clínico. Departamento de Enfermedades Cardiovasculares.
  • Maturana Barahona, Gustavo; Pontificia Universidad Católica de Chile. Hospital Clínico. Departamento de Enfermedades Cardiovasculares.
  • Zalaquett Sepúlveda, Ricardo; Pontificia Universidad Católica de Chile. Hospital Clínico. Departamento de Enfermedades Cardiovasculares.
  • Garrido O., Luis; Pontificia Universidad Católica de Chile. Hospital Clínico. Departamento de Enfermedades Cardiovasculares.
  • De la Paz P., Francisco; Pontificia Universidad Católica de Chile. Hospital Clínico. Departamento de Enfermedades Cardiovasculares.
  • Casanegra Prjnat, Pablo; Pontificia Universidad Católica de Chile. Hospital Clínico. Departamento de Enfermedades Cardiovasculares.
Rev. méd. Chile ; 127(3): 309-18, mar. 1999. tab
Article Dans Espagnol | LILACS | ID: lil-243795
ABSTRACT

Background:

Continuous improvement of dual chamber DDD pacemakers, electrode stability and programmed sequential stimulation changed the prognosis of patients implanted with these devices.

Aim:

To report our experience with the use of dual chamber pacemakers. Material and

methods:

One hundred seventy six patients (116 male), aged 13 to 91 years old, who received a dual chamber pacemaker implant, are reported. Patients were followed for a mean of 2.6 years.

Results:

Indications for DDD pacemaker were complete atrioventricular block in 43 percent, sick sinus syndrome in 32 percent, paroxysmal A-V block in 24 percent. All pacemakers were Siemens-Pacesetter and were provided with an automatic sensing and threshold device. J shaped atrium electrodes were used in 78 percent of patients and screw-in electrodes in 22 percent. Post operative complications were displacement of atrial electrode in 8 patients, of ventricular electrode in 6 patients, infection in two patients and a hematoma in one. Chronic parameters, measured after six months, were within expected ranges and allowed a good reprogramming of pacemakers. Long term programming aimed to reduce battery depletion, enhance device performance and improve hemodynamic conditions. Normal sequential stimulation was achieved in 154 patients (87 percent), 14 (85) patients died of cardiovascular disease not related to pacemaker function. Eight patients were in atrial fibrillation and were reprogrammed to VVI and DDI modes.

Conclusions:

DDD pacemakers are reliable and afford symptomatic relief in a broad spectrum of patients