Aumento da longevidade do marcapasso através de reprogramaçäo. Subutilizaçäo e benefícios / Enhancing longevity of pacemakers through reprogramming. Underutilization and cost-effectiveness
Ribeiro, Antonio Luiz Pinho; Garcia Rincón, Leonor; Oliveira, Bruna Guimaräes; Mota, Cleonice Carvalho; Pires, Marco Túlio Bacarini.
  • Ribeiro, Antonio Luiz Pinho; Universidade Federal de Minas Gerais. Faculdade de Medicina. Hospital das Clínicas. Belo Horizonte. BR
  • Garcia Rincón, Leonor; Universidade Federal de Minas Gerais. Faculdade de Medicina. Hospital das Clínicas. Belo Horizonte. BR
  • Oliveira, Bruna Guimaräes; Universidade Federal de Minas Gerais. Faculdade de Medicina. Hospital das Clínicas. Belo Horizonte. BR
  • Mota, Cleonice Carvalho; Universidade Federal de Minas Gerais. Faculdade de Medicina. Hospital das Clínicas. Belo Horizonte. BR
  • Pires, Marco Túlio Bacarini; Universidade Federal de Minas Gerais. Faculdade de Medicina. Hospital das Clínicas. Belo Horizonte. BR
Arq. bras. cardiol ; 76(6): 437-444, June 2001. ilus, tab
Article Dans Portugais, Anglais | LILACS | ID: lil-286361
ABSTRACT

OBJECTIVE:

This study was performed to observe the number of pacemakers that had never been reprogrammed after implantation, and the effect of optimised output programming on estimated longevity of pulse generators in patients with pacemaker

METHODS:

Sixty patients with Teletronics Reflex pacemakers were evaluated in a pacemaker clinic, from the time of the beginning of its activities, in June 1998, until March 1999. Telemetry was performed during the first clinic visit, and we observed how many pulse generators retained nominal output settings of the manufactures indicating the absence of reprogramming until that date. After evaluation of the capture threshold, reprogramming of pacemakers was performed with a safety margin of 2 to 2.51, and we compared the estimated longevity based on battery current at the manufacturer's settings with that based on settings achieved after reprogramming.

RESULTS:

In 95 percent of the cases, the original programmed setting was never reprogrammed before the patients attended the pacemaker clinic. Reprogramming the pacemaker prolonged estimated pulse generator life by 19.7Ý15.6 months (35.5 percent).

CONCLUSION:

The majority of the pacemakers evaluated had never been reprogrammed. Estimated pulse generator longevity can be prolonged significantly, using this simple, safe, efficacious, and cost-effective procedure
Sujets)
Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Pacemaker / Alimentations électriques / Logiciel Type d'étude: Évaluation en économique de la santé Limites du sujet: Femelle / Humains / Mâle langue: Anglais / Portugais Texte intégral: Arq. bras. cardiol Thème du journal: Cardiologie Année: 2001 Type: Article Pays d'affiliation: Brésil Institution/Pays d'affiliation: Universidade Federal de Minas Gerais/BR

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Pacemaker / Alimentations électriques / Logiciel Type d'étude: Évaluation en économique de la santé Limites du sujet: Femelle / Humains / Mâle langue: Anglais / Portugais Texte intégral: Arq. bras. cardiol Thème du journal: Cardiologie Année: 2001 Type: Article Pays d'affiliation: Brésil Institution/Pays d'affiliation: Universidade Federal de Minas Gerais/BR