Resultados preliminares do implante de lente intra-ocular fácica artisan para correção de miopia / Preliminary results of artisan phakic intraocular lens implantation to correct myopia
Akaishi, Leonardo; Martins, Herval Cavalcanti; Rezende, Rodrigo; Tzelikis, Patrick Frensel de Moraes.
  • Akaishi, Leonardo; Hospital Oftalmológico de Brasília. Brasília. BR
  • Martins, Herval Cavalcanti; Hospital Oftalmológico de Brasília. Brasília. BR
  • Rezende, Rodrigo; Hospital Oftalmológico de Brasília. Brasília. BR
  • Tzelikis, Patrick Frensel de Moraes; Hospital Oftalmológico de Brasília. Serviço de Córnea e Catarata. Brasília. BR
Arq. bras. oftalmol ; 70(3): 385-389, maio-jun. 2007. tab
Article Dans Portugais | LILACS | ID: lil-459822
RESUMO

OBJETIVOS:

Avaliar os resultados preliminares quanto à eficácia, estabilidade e segurança da lente intra-ocular de câmara anterior Artisan em olhos fácicos para correção de miopia.

MÉTODOS:

Foi analisado retrospectivamente o resultado de 34 olhos fácicos que receberam o implante de lente intra-ocular Artisan para correção de miopia. Os parâmetros avaliados foram contagem endotelial pré e pós-operatória, acuidade visual sem correção pré e pós-operatória, melhor acuidade visual com correção pré e pós-operatória, equivalente esférico pré e pós-operatório, ganho e perda de linhas de visão e complicações.

RESULTADOS:

Dezoito pacientes (34 olhos) foram incluídos neste estudo. Quatorze eram do sexo feminino (77,8 por cento) e 4 do sexo masculino. A idade média dos pacientes era de 30 anos (DP ± 7,3), variando de 21 a 46 anos. O período médio de seguimento foi de 8,5 meses (DP ± 3,6). A média de perda endotelial foi de 4,75 por cento em 6 meses. A acuidade visual sem correção pré-operatória era de 0,02 (20/800) variando de 0,01 a 0,13. O equivalente esférico médio pré-operatório na refração dinâmica era de -13,25 D, variando de -5,75 D a -19,75 D. No último seguimento a acuidade visual sem correção era de 0,64 (20/32) variando de 0,33 a 1,00. O equivalente esférico médio na refração dinâmica era de -1,18 D (DP ± 0,92) variando de + 0,25 a -3,0. Entre as complicações observadas em nosso estudo 1 (3,4 por cento) paciente apresentou deslocamento da lente por trauma necessitando de uma segunda intervenção para reposicionamento.

CONCLUSÃO:

O uso de lente fácica de câmara anterior Artisan para correção de miopia no presente estudo se mostrou seguro, eficaz e com boa previsibilidade. No entanto, estudo prospectivo com maior número de casos e maior seguimento é necessário para determinar a segurança do procedimento a longo prazo.
ABSTRACT

PURPOSE:

To observe the preliminary results based on the efficacy, stability and safety of Artisan anterior chamber intra-ocular lens for the correction of myopia in the first 6 months.

METHODS:

A retrospective study of 18 patients (34 eyes) that received the implantation of Artisan anterior chamber intraocular lens for the correction of myopia was conducted. Best visual preoperative acuity, best visual postoperative acuity, dynamic preoperative and postoperative refraction, loss and improvement of lines of vision and complications were evaluated.

RESULTS:

Eighteen patients (34 eyes) were included in this study. Fourteen patients were females (77.8 percent) and 4 were males. The mean age was 30 years (SD ± 7.3), range from 21 to 46 years. They were followed up for an average of 8.5 months (SD ± 3.6). The mean endothelial cell loss was 4.75 percent at 6 months follow-up. The average preoperative uncorrected visual acuity was 0.02 (20/800) range from 0.01 to 0.13. The average preoperative spherical equivalent was -13.25 D, range from -5.75 D to -19,75 D. On the last follow-up, uncorrected visual acuity was 0.64 (20/32), range from 0.33 to 1.00; the average spherical equivalent was -1.18 D (DP ± 0.92) range from +0.25 to -3.0. Complications were observed in 1 case (3.4 percent) of our study, in one patient who had a dislocation of the lens after an ocular trauma.

CONCLUSION:

The use of Artisan anterior chamber intraocular lens in phakic eyes for the correction of myopia was safe, effective and predictable. However, a larger prospective study with a higher number of cases and longer follow-up is necessary to determine long-term safety of the lens.
Sujets)

Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Pose d'implant intraoculaire / Lentilles intraoculaires / Myopie Type d'étude: Étude observationnelle / Étude pronostique / Facteurs de risque Limites du sujet: Adulte / Femelle / Humains / Mâle langue: Portugais Texte intégral: Arq. bras. oftalmol Thème du journal: Ophtalmologie Année: 2007 Type: Article Pays d'affiliation: Brésil Institution/Pays d'affiliation: Hospital Oftalmológico de Brasília/BR

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Pose d'implant intraoculaire / Lentilles intraoculaires / Myopie Type d'étude: Étude observationnelle / Étude pronostique / Facteurs de risque Limites du sujet: Adulte / Femelle / Humains / Mâle langue: Portugais Texte intégral: Arq. bras. oftalmol Thème du journal: Ophtalmologie Année: 2007 Type: Article Pays d'affiliation: Brésil Institution/Pays d'affiliation: Hospital Oftalmológico de Brasília/BR