Your browser doesn't support javascript.
loading
Resultados preliminares de um sistema de topografia de grande-ângulo usando videoceratógrafo com discos de Plácido / Preliminary results of a wide-angle corneal topography system for Placido-based videokeratographers
Carvalho, Luis Alberto Vieira de.
Affiliation
  • Carvalho, Luis Alberto Vieira de; Universidade de São Paulo. Instituto de Física de São Carlos. Grupo de Óptica. Laboratório de Óptica Oftálmica. São Carlos. BR
Arq. bras. oftalmol ; Arq. bras. oftalmol;71(2): 197-206, mar.-abr. 2008. ilus
Article de Pt | LILACS | ID: lil-483027
Bibliothèque responsable: BR1.1
RESUMO

OBJETIVO:

Desenvolver a instrumentação e o "software" para topografia de córnea de grande-ângulo usando o tradicional disco de Plácido. O objetivo é permitir o mapeamento de uma região maior da córnea para topógrafos de córnea que usem a técnica de Plácido, fazendo-se uma adaptação simples na mira.

MÉTODOS:

Utilizando o tradicional disco de Plácido de um topógrafo de córnea tradicional, 9 LEDs (Light Emitting Diodes) foram adaptados no anteparo cônico para que o paciente voluntário pudesse fixar o olhar em diferentes direções. Para cada direção imagens de Plácido foram digitalizadas e processadas para formar, por meio de algoritmo envolvendo elementos sofisticados de computação gráfica, um mapa tridimensional completo da córnea toda.

RESULTADOS:

Resultados apresentados neste trabalho mostram que uma região de até 100 por cento maior pode ser mapeada usando esta técnica, permitindo que o clínico mapeie até próximo ao limbo da córnea. São apresentados aqui os resultados para uma superfície esférica de calibração e também para uma córnea in vivo com alto grau de astigmatismo, mostrando a curvatura e elevação.

CONCLUSÃO:

Acredita-se que esta nova técnica pode propiciar a melhoria de alguns processos, como por exemplo adaptação de lentes de contato, algoritmos para ablações costumizadas para hipermetropia, entre outros.
ABSTRACT

PURPOSE:

To develop the instrumentation and software for wide-angle corneal topography using a Placido-based videokeratographer. The objective is to allow the measurement of a greater area of the cornea using a simple adaptation to the Placido mire.

METHODS:

Using the traditional Placido disc of a commercial corneal topographer, 9 LEDs (Light Emitting Diodes) were installed on the conic mire so that the voluntary patient could gaze in different directions. For each direction Placido images were acquired and processed to form, using a sophisticated computer graphics algorithm, a tridimensional map of the whole cornea.

RESULTS:

It was shown that up to 100 percent more area of the cornea may be mapped using the here described technique. We present results for a spherical surface calibration and also for a highly astigmatic cornea and analyze quantitatively the additional area that is recovered in terms of curvature and true elevation.

CONCLUSIONS:

We believe that this new technique opens opportunities for the improvement of other applications which might benefit from this additional information, such as contact lens adaptation and design improvement, algorithms for customized refractive surgery, among others.
Sujet(s)
Mots clés
Texte intégral: 1 Indice: LILACS Sujet Principal: Traitement d'image par ordinateur / Logiciel / Éclairage / Cornée / Topographie cornéenne Type d'étude: Diagnostic_studies Limites du sujet: Humans langue: Pt Texte intégral: Arq. bras. oftalmol Thème du journal: OFTALMOLOGIA Année: 2008 Type: Article
Texte intégral: 1 Indice: LILACS Sujet Principal: Traitement d'image par ordinateur / Logiciel / Éclairage / Cornée / Topographie cornéenne Type d'étude: Diagnostic_studies Limites du sujet: Humans langue: Pt Texte intégral: Arq. bras. oftalmol Thème du journal: OFTALMOLOGIA Année: 2008 Type: Article