Avaliação do kerma no ar na superfície de entrada e da qualidade da imagem em radiografias de tórax / Evaluation of entrance surface air kerma rate and clinical images quality in chest radiography
Souza, Angelo Bernardo Brasil de; Dias, Simone Kodlulovich; Augusto, Fernando Mecca; Marques, Winston Andrade.
  • Souza, Angelo Bernardo Brasil de; Universidade Federal do Rio de Janeiro. Instituto Alberto Luiz Coimbra de Pós-Graduação e Pesquisa em Engenharia. Programa de Engenharia Biomédica. Rio de Janeiro. BR
  • Dias, Simone Kodlulovich; Instituto de Radioproteção e Dosimetria. Comissão Nacional de Energia Nuclear. Rio de Janeiro. BR
  • Augusto, Fernando Mecca; Instituto Nacional de Câncer. Setor de Radiodiagnóstico da Física médica. Rio de Janeiro. BR
  • Marques, Winston Andrade; Instituto Nacional de Câncer. Rio de Janeiro. BR
Radiol. bras ; 42(5): 315-319, set.-out. 2009. graf, tab
Article Dans Portugais | LILACS | ID: lil-530180
RESUMO

OBJETIVO:

Avaliar os fatores de técnica utilizados durante exames de tórax póstero-anterior em pacientes-padrão e correlacioná-los aos valores de kerma no ar na superfície de entrada e à qualidade das imagens clínicas.

MATERIAIS E MÉTODOS:

O estudo foi realizado em três hospitais do Rio de Janeiro, num total de cinco salas de raios X, com dez pacientes por sala. As imagens foram avaliadas pelos radiologistas dos serviços segundo o protocolo europeu. O kerma no ar na superfície de entrada foi estimado a partir da curva de rendimento do equipamento de raios X, que foi obtida utilizando câmara de ionização acoplada a um eletrômetro. Análise de variância foi realizada para verificar se a diferença entre os valores de kerma no ar na superfície de entrada é significativa.

RESULTADOS:

Os valores de kerma no ar na superfície de entrada variaram entre 0,05 e 0,26 mGy, com média 60 por cento inferior ao nível de referência publicado na Portaria 453. Das imagens avaliadas, 98 por cento atenderam acima de 65 por cento dos critérios de qualidade.

CONCLUSÃO:

Para um padrão de qualidade da imagem, aceitável para o diagnóstico, verificou-se ampla variação do kerma no ar na superfície de entrada para pacientes-padrão. Isto demonstra a falta de padronização dos fatores de técnica e a existência de um potencial de redução do valor do kerma no ar na superfície de entrada.
ABSTRACT

OBJECTIVE:

To evaluate technical factors involved in posteroanterior chest radiography in standard patients, and correlating them with entrance surface air kerma rate and with the clinical images quality. MATERIALS AND

METHODS:

The present study was developed at three hospitals in Rio de Janeiro, Brazil, involving a total of five x-ray rooms, with ten patients per room. The images were evaluated by the radiologists of the institutions, in accordance with the European protocol. The estimation of surface entrance air kerma rate was based on the x-ray equipment output curve obtained with an ionization chamber coupled with an electrometer. Variance analysis was performed to evaluate the significance of the difference between entrance surface air kerma rates.

RESULTS:

Entrance surface air kerma rates ranged between 0.05 and 0.26 mGy, with a mean value 60 percent lower than the reference level established by the Order (Portaria) 453. Among the images evaluated, 98 percent fulfilled > 65 percent of the images quality criteria.

CONCLUSION:

For a standard image quality that is acceptable for diagnosis purposes, a significant variation was observed in the entrance surface air kerma rate for standard patients, demonstrating the lack of technical factors standardization and a potential for decreasing the entrance surface air kerma rate.
Sujets)

Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Radiographie thoracique / Kerma Type d'étude: Guide de pratique langue: Portugais Texte intégral: Radiol. bras Thème du journal: Imagerie diagnostique / Radiologie Année: 2009 Type: Article Pays d'affiliation: Brésil Institution/Pays d'affiliation: Instituto Nacional de Câncer/BR / Instituto de Radioproteção e Dosimetria/BR / Universidade Federal do Rio de Janeiro/BR

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Radiographie thoracique / Kerma Type d'étude: Guide de pratique langue: Portugais Texte intégral: Radiol. bras Thème du journal: Imagerie diagnostique / Radiologie Année: 2009 Type: Article Pays d'affiliation: Brésil Institution/Pays d'affiliation: Instituto Nacional de Câncer/BR / Instituto de Radioproteção e Dosimetria/BR / Universidade Federal do Rio de Janeiro/BR