Your browser doesn't support javascript.
loading
Estudo comparativo da qualidade da imagem e do kerma, de entrada e de saída, em simulador de tórax utilizando sistemas analógico e digitalizado CR de aquisição de imagens / Comparative study of image quality and entrance and exit air kerma measurements on chest phantom utilizing analog and CR digital imaging systems
Luz, Renata Matos da; Hoff, Gabriela.
  • Luz, Renata Matos da; s.af
  • Hoff, Gabriela; Pontifícia Universidade Católica do Rio Grande do Sul. Porto Alegre. BR
Radiol. bras ; 43(1): 39-45, jan.-fev. 2010. tab, ilus
Article Dans Portugais | LILACS | ID: lil-542688
RESUMO

OBJETIVO:

O processo de migração de sistemas analógicos para digitalizados, para aplicações diagnósticas, requer cuidados específicos, a fim de manter a qualidade das imagens e minimizar a dose no paciente. Este trabalho tem como objetivo analisar e comparar a qualidade da imagem e o kerma num simulador não antropomórfico de tórax gerados por sistemas analógicos e digitalizados CR.

MATERIAIS E MÉTODOS:

Foram analisados a qualidade da imagem e o kerma, de entrada e saída, no simulador para dois equipamentos de raios X diferentes (Siemens e Emic), com diferentes sistemas de retificação de onda (12 pulsos e alta frequência). Ambos os sistemas (analógico e digitalizado) estavam sendo utilizados no mesmo local. Foram geradas imagens em filme e em image plates.

RESULTADOS:

Foi observado aumento na tensão e/ou na carga transportada pelo tubo de raios X quando houve a migração para o sistema CR, para manutenção das características diagnósticas da imagem. Isto resultou em aumento de kerma coletado.

CONCLUSÃO:

As maiores diferenças determinadas (aumento de dose e redução da qualidade da imagem) foram observadas no equipamento com retificação de onda de 12 pulsos e transdutor de imagem CR (image plates).
ABSTRACT

OBJECTIVE:

The process of migration from analog to digital imaging system requires specific attention to preserve images quality and minimizing the dose to the patients. The present study was aimed at analyzing and comparing images quality and entrance and exist air kerma measurements in a non-anthropomorphic chest phantom with analog and CR digital imaging systems. MATERIALS AND

METHODS:

Two different X-ray units (Siemens and Emic) with different wave rectification systems (12-pulse and high-frequency) were utilized in a single institution along the process of migration from analog to CR digital imaging system. Images were acquired on films and image plates.

RESULTS:

An increase on the X-ray tube peak voltage and/or load was observed in the migration from analog to CR digital system to preserve the diagnostic quality of the image. As a result, an increase in the air kerma rate was observed.

CONCLUSION:

The greatest differences (increase in dose and decrease in image quality) were observed with the 12-pulse wave rectification system with the CR image transducer (imaging plates).
Sujets)

Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Conversion analogique-numérique / Amélioration d'image radiographique / Technologie radiologique / Kerma langue: Portugais Texte intégral: Radiol. bras Thème du journal: Imagerie diagnostique / Radiologie Année: 2010 Type: Article Pays d'affiliation: Brésil Institution/Pays d'affiliation: Pontifícia Universidade Católica do Rio Grande do Sul/BR

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Conversion analogique-numérique / Amélioration d'image radiographique / Technologie radiologique / Kerma langue: Portugais Texte intégral: Radiol. bras Thème du journal: Imagerie diagnostique / Radiologie Année: 2010 Type: Article Pays d'affiliation: Brésil Institution/Pays d'affiliation: Pontifícia Universidade Católica do Rio Grande do Sul/BR