Your browser doesn't support javascript.
loading
Cierre percutáneo de la comunicación inter-atrial con dispositivo de Amplatzer: experiencia de 42 casos / Percutaneous closure of inter-atrial communication with the Amplatzer device: experience with 42 cases
Munayer-Calderón, Jaime E; Aldana-Pérez, Tomás; Carpio-Hernández, Juan Carlos; Lázaro-Castillo, José Luis; Ramírez-Pérez, Homero; San-Luis-Miranda, Raúl; Pérez-Montiel, Inocencio.
  • Munayer-Calderón, Jaime E; Instituto Mexicano del Seguro Social. Centro Médico Nacional La Raza. Hospital General. Departamento de Hemodinamia. México, D. F. MX
  • Aldana-Pérez, Tomás; Instituto Mexicano del Seguro Social. Centro Médico Nacional La Raza. Hospital General. Departamento de Hemodinamia. México, D. F. MX
  • Carpio-Hernández, Juan Carlos; Instituto Mexicano del Seguro Social. Centro Médico Nacional La Raza. Hospital General. Departamento de Cardiopediatría. México, D. F. MX
  • Lázaro-Castillo, José Luis; Instituto Mexicano del Seguro Social. Centro Médico Nacional La Raza. Hospital General. Departamento de Hemodinamia. México, D. F. MX
  • Ramírez-Pérez, Homero; Instituto Mexicano del Seguro Social. Centro Médico Nacional La Raza. Hospital General. Departamento de Hemodinamia. México, D. F. MX
  • San-Luis-Miranda, Raúl; Instituto Mexicano del Seguro Social. Centro Médico Nacional La Raza. Hospital General. Departamento de Hemodinamia. México, D. F. MX
  • Pérez-Montiel, Inocencio; Instituto Mexicano del Seguro Social. Centro Médico Nacional La Raza. México, D. F. MX
Arch. cardiol. Méx ; 79(2): 104-106, abr.-jun. 2009.
Article Dans Espagnol | LILACS | ID: lil-565725
ABSTRACT
We report our experience on 42 patients treated with atrial septal defect (ASD) occlusion using Amplatzer device. Thirty patients were females and 12 were males, mean ages 26.5-years-old +/- 12 years (interval from 7 to 69 years). Average weight was 57.1 +/- 13.8 kg (32.3-85.0 kg). Forty-two devices (ranging diameters from 13 mm to 40 mm) were deployed, 40 of which were placed successfully and attempts were unsuccessful in two cases (diameter devices 36 mm and 40 mm, respectively). We used the [quot ]balloon on the left or right upper pulmonary vein[quot ] in 5 patients, achieving good deployment. Echocardiography showed total occlusion in 37 patients (93.5%), trivial leak in 2 (4.7%), and light leak in 1 patient (2.3%). Follow up was at 1 to 12 months (mean 6.5). Total occlusion was observed at one month on both patients with trivial leak, and at 6 months on the patient with light leak. Failure to deploy the device appropriately on the two patients with unsuccessful result was due to unfavorable anatomy very large defects in both cases (occluder size 36 mm and 40 mm), very thin postero-superior 6 mm rim on one of them and aortic rim absence on the other one. Stretched diameters were 34 mm and 38 mm on patients with 36 mm and 40 mm devices, respectively. Both of them were sent to surgery. We conclude that percutaneus closure of atrial septal defect with the Amplatzer device is a save and have good results.
Sujets)
Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Prothèses et implants / Communications interauriculaires Limites du sujet: Adolescent / Adulte / Adulte très âgé / Enfant / Femelle / Humains / Mâle langue: Espagnol Texte intégral: Arch. cardiol. Méx Thème du journal: Cardiologie Année: 2009 Type: Article Pays d'affiliation: Mexique Institution/Pays d'affiliation: Instituto Mexicano del Seguro Social/MX

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: LILAS (Amériques) Sujet Principal: Prothèses et implants / Communications interauriculaires Limites du sujet: Adolescent / Adulte / Adulte très âgé / Enfant / Femelle / Humains / Mâle langue: Espagnol Texte intégral: Arch. cardiol. Méx Thème du journal: Cardiologie Année: 2009 Type: Article Pays d'affiliation: Mexique Institution/Pays d'affiliation: Instituto Mexicano del Seguro Social/MX