Análise biomecânica de três tipos de conexões protéticas em implantes dentários de diâmetro reduzido / Biomechanical analysis of three types of prosthetic connections on reduced diameter dental implants
ImplantNews
; 11(5): 595-599, 2014. tab
Article
de Pt
| LILACS, BBO
| ID: lil-731522
Bibliothèque responsable:
BR243.1
RESUMO
Objetivos:
avaliar o comportamento biomecânico de três tipos de conexão protética em implantes de diâmetro reduzido. Material eMétodos:
implantes com diâmetro 3,25 mm (plataforma 3,4 mm) foram inseridos em polioximetileno (poliacetal) com torque máximo de 45 Ncm e receberam componentes protéticos calcináveis divididos em três grupos G1 cone-morse (CM), hexágono interno (HI), e Hexágono externo (HE). Todo os pilares foram sobrefundidos em liga de CoCr e aparafusados com torque de 20 Ncm conforme recomendações do fabricante. A avaliação biomecânica ocorreu conforme a norma ISO 14801-2007 carregamento compressivo 30º e fadiga cíclica (f=15Hz, 5 milhões de ciclos) em ambiente seco. A análise estatística foi conduzida usando-se os testes Anova e Tukey (nível de significância 5%).Resultados:
os valores compressivos médios foram G1 (317,43 N), G2 (559,44 N), e G3 (290,83 N). Diferenças signifi cativas foram vistas entre G1 e G2 (p < 0,001) e entre G2 e G3 (p < 0,0001). Não houve diferença entre grupos no teste de fadiga. Os valores correspondentes foram G1 (190,80 N), G2 (163,25 N), e G3 (141,77 N).Conclusões:
todos os grupos demonstraram desempenho biomecânico aceitávelABSTRACT
Objectives:
to evaluate the biomechanical behavior of three types of prosthetic abutment connections in reduced implant diameters. Material andMethods:
3.25 mm-diameter dental implants (3.4 mm implant platform) were inserted in polyoxymethylene (polyacetal) resin with maximum torque of 45Ncm and received burnout prosthetic abutments divided into three groups G1 cone-morse (CM), Internal Hex (IH), and External Hex (EH). All abutments were overcast in a CoCr alloy and fastened to 20 Ncm following manufacturers instructions. Biomechanical evaluation was performed according to ISO 14801-2007 standards 30º compressive and fatigue loading (f = 15 Hz, 5 million cycles) in dry environment. Statistical analysis was conducted with Anova and post-hoc Tukey tests (5% signifi cance level).Results:
mean compressive values were G1 (317.43 N), G2 (559.44 N), and G3 (290.83 N). Signifi cant differences were seen between G1 and G2 (p < 0.001) and between G2 and G3 (p < 0.0001). No differences among groups were seen for fatigue loading. The correspondent endurance limits were G1 (190.80 N), G2 (163.25 N), and G3 (141.77 N).Conclusions:
all groups demonstrated acceptable biomechanical performance.Mots clés