Acurácia e precisão da angiografia coronária quantitativa online com calibração automática: um estudo piloto / Accuracy and precision of online quantitative coronary angiography with automatic calibration: a pilot study
Rev. bras. cardiol. invasiva
; 23(1): 58-60, abr.-jun.2015. tab
Article
de Pt
| LILACS
| ID: lil-782177
Bibliothèque responsable:
BR44.1
RESUMO
O objetivo deste estudo foi avaliar o grau de acurácia, precisão, correlação e concordância entre as medidas realizadas por um software de angiografia coronária quantitativa (QCA) online de calibração automática. Métodos:
Estudo piloto que analisou imagens angiográficas de dez pacientes com um software de QCA online nas calibrações automáticas Auto ISO (calibração automática do isocentro) e Auto TOD (Table-to Object Distance). Foi realizada a medida do calibre do cateter pelos dois métodos, e o diâmetro de referência foi computado. Essas medidas foram comparadas com a medida do diâmetro do cateter quanto à acurácia, precisão e concordância.Resultados:
O diâmetro médio real dos cateteres era de 1,75 ± 0,32 mm (variando entre 1,33 e 2,67 mm). A aferição dos cateteres pela QCA TOD e pela QCA ISO resultou em diâmetros médios de 1,78 ± 0,37 mm e 1,88 ± 0,38 mm, respectivamente. A acurácia/precisão da QCA TOD e da QCA ISO foi 0,03 m/0,21 mm e 0,12 mm/0,20 mm, respectivamente. As medidas da QCA TOD e da QCA ISO estiveram entre os limites de concordância em 96,3 e 94,7% dos casos, respectivamente. As medidas da QCA TOD e da QCA ISO correlacionaram-se significativamente (rs = 0,93; p < 0,01). No entanto, apesar da pequena diferença entre os métodos (0,10 ± 0,10 mm), as medidas da QCA ISO foram estatisticamente maiores que aquelas obtidas com a QCA TOD (p < 0,01).Conclusões:
A QCA online com calibração automática apresentou boa acurácia, precisão e correlação,podendo representar uma ferramenta promissora no laboratório de hemodinâmica...ABSTRACT
The objective of this study was to evaluate the degree of accuracy, precision, correlation, andagreement between the measurements performed by online Quantitative Coronary Angiography (QCA)software with automatic calibration.Methods:
Pilot study that analyzed angiographic images of ten patients through online QCA softwareusing Auto ISO (automatic calibration isocenter) and Auto TOD (Table-to-Object Distance) automatic calibration. Catheter size was measured by both methods and the reference diameter was computed. These measurements were compared with the measurement of catheter diameter regarding accuracy, precision,and agreement.Results:
The actual average of the catheter diameter was 1.75 ± 0.32 mm (range 1.33 to 2.67 mm). The measurement of catheters by TOD QCA and ISO QCA resulted in mean diameters of 1.78 ± 0.37 mm and 1.88± 0.38 mm, respectively. The accuracy/precision of the TOD QCA and the ISO QCA was 0.03 mm/0.21 mmand 0.12 mm/0.20 mm, respectively. The TOD QCA and ISO QCA measures were among the limits ofagreement in 96.3 and 94.7% of cases, respectively, and were significantly correlated (rs = 0.93, p < 0.01). However, despite the small difference between the methods (0.10 ± 0.10 mm), the ISO QCA measures were significantly higher than those obtained by the TOD QCA (p < 0.01).Conclusions:
Online QCA with automatic calibration has good accuracy, precision, and correlation, which may represent a promising tool in the catheterization laboratory...
Texte intégral:
1
Indice:
LILACS
Sujet Principal:
Coronarographie
/
Maladie coronarienne
/
Précision de la mesure dimensionnelle
Type d'étude:
Prognostic_studies
Limites du sujet:
Female
/
Humans
/
Male
langue:
Pt
Texte intégral:
Rev. bras. cardiol. invasiva
Thème du journal:
CARDIOLOGIA
/
CIRURGIA GERAL
Année:
2015
Type:
Article