Avaliação da resistência imposta pelos filtros trocadores de calor e umidade às vias aéreas de pacientes com suporte ventilatório mecânico / Evaluation of resistance imposed by heat exchangers and moisture filters of airway in patients with mechanical ventilatory support
Fisioter. Bras
;
12(5): 353-358, set.-out. 2011. ilus, tab
Article
Dans Portugais
| LILACS
| ID: lil-785329
RESUMO
Introdução:
Os FTCUs (filtros trocadores de calor e umidade)são dispositivos que combinam propriedades de umidificação e retenção bacteriana através de membranas que protegem pacientes mecanicamente ventilados.Objetivo:
Analisar, em um modelo experimental,os efeitos de diferentes marcas de FTCU, em diferentes tempos de uso sobre resistência das vias aéreas de pacientes com suporte ventilatório artificial. Material emétodo:
Trata-se de um ensaio clínico randomizado, a amostra constitui-se de 30 filtros de três diferentes marcas, utilizados por 48 horas em pacientes internados numa unidade de terapia intensiva. A variável estudada consistiu na resistência ao fluxo de ar, analisada em 0, 24 e 48 horas para cada marca de FTCU.Resultados:
Não houve diferença estatisticamente significativa na resistência ao fluxo de ar entre as marcas no tempo de uso 24 horas (p = 0,489) e 48 horas (p = 0,374), apenas no momento0 hora de utilização houve uma diferença estatisticamente significativa(p = 0,027) entre as marcas estudadas, e, além disso, dentro da mesma marca não houve diferença estatística na resistência ao fluxo de ar com o tempo de uso.Conclusão:
A utilização por 48 horas dos FTCUs não interfere na resistência ao fluxo de ar de pacientes mecanicamente ventilados promovendo segurança no seu uso.ABSTRACT
Introduction:
The FTCUs (heat and moisture exchange filters)are devices that combine wetting properties and retention through bacterial membranes that protect mechanically ventilated patients.Objective:
To analyze, in an experimental model, the effects ofdifferent brands of FTCU at different times of use on the airways of patients with mechanical ventilation. Material andmethods:
Randomized clinical trial with a sample of 30 filters used threedifferent brands for 48 hours in ICU patients. The variable studied was the resistance to airflow analyzed at 0, 24 and 48 hours for each brand of FTCU.Results:
There were no statistically significant difference in resistance to air flow between the brands used 24 hours(p = 0.489) and 48 hours (p = 0.374). Just in time 0 hour of usethere was a statistically significant difference (p = 0.027) among thebrands studied. Within the same brand no statistical difference wasobserved in the resistance to air flow with time of use.Conclusion:
The use of filters does not interfere with FTCUs resistance to airflowin mechanically ventilated patients promoting security in its use.
Recherche sur Google
Indice:
LILAS (Amériques)
Sujet Principal:
Ventilation artificielle
/
Filtres
/
Résistance des voies aériennes
/
Unités de soins intensifs
Type d'étude:
Essai clinique contrôlé
/
Étude pronostique
Limites du sujet:
Humains
langue:
Portugais
Texte intégral:
Fisioter. Bras
Thème du journal:
MEDICINA FISICA E REABILITACAO
Année:
2011
Type:
Article
Pays d'affiliation:
Brésil
Institution/Pays d'affiliation:
UFSC+BR
/
Hospital Universitário HU-UFSC/BR
/
IEB-UFSC/BR
/
UFSC+BR
/
UDESC/BR
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS